电流放大器 Fluke 5220A
后面板
厂家: 美国福禄克
尺寸: 18x43x56cm
Transconductance Amplifier
详细指标 pdf文档
- Transconductance: 1 siemens (1 A per volt)
- Output range: 0 to 20 A dc or rms ac
- DC accuracy: ±(0.025% of output + 1 mA)
- AC accuracy: ±(0.05% of output + 1 mA) from 30 Hz to 1 kHz, and ±(0.05%
of output + 1 mA) x f from 1 kHz to 5 kHz, where f is frequency in kHz
- Short-term stability dc: output changes less than ±(0.005% + 200
µA) in 10 minutes, with constant line, load, and temperature
- Short-term stability ac: output changes less than ±(0.01% + 500
µA) in 10 minutes, with constant line, load, and temperature
- 100, 110, 115, 120, 200, 220, 230, 240 Vac +10%; 50-60 Hz operation
- Weight: 50 lb
The 5220A Transconductance Amplifier allows for the
calibration of AC or DC meters and shunts and current functions of digital
multimeters that measure up to 20 A. A known input voltage of 1 to 20 volts
produces a known output current of 1 to 20 amps. The 5220A is designed to be
controlled by the 5700A or 5100B series calibrators but may be driven by any
other voltage source.
咨询电话: 132- 4173- 5460 , 邮箱: 52634729@QQ.com