Hewlett Packard 3577B is a high-performance, yet economical, two-channel, (optional three-channel) network analyzer for use in both research and development and manufacturing. It is used to measure magnitude/phase, insertion loss, group delay, SWR, electrical length, and gain compression from 5 Hz to 200 MHz. When used with the 35676A/B reflection/transmission kit, it can also measure return loss, reflection coefficient, and impedance.
With the optional HP Instrument BASIC, the 3577B can execute user-written programs designed to automate measurement systems, compute parameters (such as pass band ripple and 3 dB bandwidth) or customize the user-interface. This includes system control of other HP-IB instruments and peripherals via the HP-IB. A programmable I/O port, located on the rear panel, extends the HP Instrument BASIC control to non-HP-IB devices, such as device handlers, environmental chambers, and even the device-under-test itself.
Of course, the 3577B is fully programmable, either internally with HP Instrument BASIC, or externally via the Hewlett-Packard Interface Bus (HP-IB).
An optional high-stability frequency reference oven is available for those users not having an in-house frequency standard. HP also offers a wide range of accessories, including S-parameter and reflection/transmission test sets, as well as signal injection devices for feedback control systems such as power supplies and servo loops.
Unprecedented Measurement Precision
This network analyzer has the accuracy and resolution required to characterize the most demanding narrowband devices and the flexibility to quickly characterize wideband devices. Dynamic magnitude and phase accuracy are 0.02 dB and 0.2 degree, respectively. Device response can be examined in fine detail with 0.001 dB, 0.005 degree, and 1 ps resolution. A built-in synthesized LO and tracking generator provide superb frequency accuracy with 0.001 Hz resolution. The 100 dB plus dynamic range and -130 dBm noise floor meet the needs of the most demanding measurements.
Built-in three-term error correction removes errors due to directivity, frequency response, and source mismatch for one-port measurements. Similarly, vector normalization enhances the accuracy of two-port measurements.
High-Throughput for Manufacturing
The 3577B brings high-throughput network analysis to manufacturing without compromising accuracy. Take advantage of features such as discrete sweep, comparison to limit lines, and PASS/FAIL testing to speed your manufacturing test of crystals, filters, and other devices.
5Hz-200MHz Network Analyzer Specifications |
SOURCE | |||
Frequency | |||
Range | 5 Hz to 200 MHz | ||
Resolution | 0.001 Hz | ||
Amplitude | |||
Range | + 15 dBm to -49 dBm (1.26 V rms to 793µ V rms: 2 dB V to -62 dBV) into a 50 Ω load | ||
Resolution | 0.1 dB | ||
Accuracy | ± 1 dB at + 15 dBm and 100 kHz. Below +15 dBm, add the greater of ± 0.02 dB/dB or 0.2 dB | ||
Flatness | 1.5 dB p-p from 5 Hz to 200 MHz | ||
Impedance | 50 Ω; > 20 dB return loss at all levels | ||
RF Output Connector | 50 Ω Type N female | ||
Sweep Types | Linear, discrete, alternate, cw and log frequency; log amplitude | ||
Sweep Time | 100 ms/span to 6553 sec/span for frequency sweep; 1 ms/step to 16 s/step for amplitude sweep | ||
Sweep Modes | Continuous, single, manual | ||
Trigger Modes | Free run, immediate, line, external | ||
RECEIVER | |||
Input Characteristics | |||
Frequency Range | 5 Hz to 200 MHz | ||
Inputs | Two receiver inputs (A,R); third receiver input (B) is optional | ||
Input Impedance | Selectable 50 Ω with > 25 dB return loss, or 1 MΩ in parallel with approximately 30 pF | ||
Input Connectors | 50 Ω Type N female | ||
Resolution Bandwidth | Selectable 1 kHz, 100 Hz, 10 Hz, or 1 Hz | ||
Magnitude Characteristics | |||
Range | Full scale input to sensitivity | ||
Display Units | dB, dBm, dBV, V, and linear ratio | ||
Accuracy (at 100 kHz, 25°C, and full scale input) |
Absolute (A, B, R): ±
0.2 dB Ratio (A/R, B/R, A/B): ± 0.15 dB (50 Ω); ± 0.2 dB (1 MΩ) |
||
Dynamic Accuracy | Error |
Input level relative to full scale input |
|
Resolution Bandwidth | |||
1 kHz, 100 Hz, 10 Hz | 1 Hz | ||
± .04 dB ± .02 dB ± .05 dB ± .15 dB ± .75 dB ± .75 dB |
± .04 dB ± .02 dB ± .05 dB ± .25 dB ± .75 dB ± 3.00 dB |
0 dB to -10 dB -10 dB to -50 dB -50 dB to -60 dB -60 dB to -80 dB -80 dB to -90 dB -90 dB to -100 dB |
|
Frequency Response (when driven from a 50 Ω source and with 50 Ω receiver input impedance) | |||
Absolute (A, B, R) | 0.3 dB p-p from 20 Hz to 20 MHz; 0.6 dB p-p from 5 Hz to 200 MHz | ||
Ratio (A/R, B/R, A/B) | 0.3 dB p-p from 20 Hz to 20 MHz; 0.4 dB from 5 Hz to 200 MHz | ||
Stability | |||
Temperature | Typically < ± 0.02 dB/°C | ||
Time | Typically ± 0.05 dB/hour at 25° C | ||
Phase Characteristics (A/R, B/R, A/B) | |||
Range | ±180° | ||
Accuracy | At 100 kHz, 25°C, and Full Scale Input: ±2.0° | ||
Dynamic Accuracy |
Error |
Input Level Relative to Full Scale Input |
|
± .4 degree |
0 dB to -10 dB -10 dB to -50 dB -50 dB to -60 dB -60 dB to -80 dB -80 dB to -100 dB |
||
Temperature Stability | Typically < ± 0.05°C | ||
Time Stability | Typically ± 0.05° hr at 25°C | ||
Delay Characteristics | |||
Range | 1 ps to 20,000s | ||
Resolution | 0.01 ns/div to 1000s/div | ||
Limit Test | Twenty segments for each trace per limit test. Delay between sweeps approximately 10 to 120 ms. | ||
GENERAL CHARACTERISTICS | |||
Traces | |||
Number of Traces | Two simultaneous traces may be present with a rectangular graticule. One trace with Polar or Smith graticules. | ||
Markers | Each trace has one main marker and an offset marker. | ||
Reference Line Position | |||
Rectangular Graticule | 0% to 100% full scale deflection in 0.05% increments. Logarithmic and linear. | ||
Polar/Smith Chart Graticule | ± 500° in 0.001° increments. | ||
Limit Test | Twenty segments for each trace per limit test. Added delay between sweeps approximately 10 to 120 ms. | ||
Noise Averaging | |||
Type | Exponentially weighted vector averaging on successive sweep data. | ||
Averaging Factor | Selectable 1 (off), 4, 8, 16, 32, 64, 128, 256 | ||
Linear Phase Slope Compensation | Provides linear phase slope offset of -72,000°/span to +72,000°/span. | ||
Calibration | |||
Transmission | Both traces can be normalized to measured data with full accuracy and resolution. | ||
Reflection | Corrects for directivity, frequency response, and source match errors | ||
Programming | |||
Remote Programming | Via the Hewlett-Packard Interface Bus (HP-IB). the HP 35677A/B S-parameter test sets are programmable through the HP 3577A interface only. | ||
Plotter Control | Directly compatible with HP-IB graphics plotters that use Hewlett-Packard Graphics Language (HP-GL) with listen-only capability. | ||
Save/recall | Front-panel setups can be stored in non-volatile memory locations 1 through 5. Last state is saved when power is removed. | ||
Operating Conditions | |||
Power | 115 V + 10%, -25% (47 Hz to 440 Hz), or 230 V + 10%, -15% (47 Hz to 66 Hz), 450 VA maximum | ||
Weight | Net, 31 kg ; shipping, 41 kg | ||
Size | 222 X 426 X 578 mm |