9354AL 9354TM
数字存储示波器 LeCroy 935x系列
厂家: 美国力科
尺寸: 21x37x41cm
重量: 13kg
主要指标
- 500M带宽,4通道
- 采样率最高可达2 GS/s
- 9354AL可达8M存储深度,9354TM可达2M存储深度
- Innovative Peak Detect Mode
- Fully programmable via GPIB and RS-232-C
- Automatic PASS/FAIL testing
- Advanced Signal Processing
- Internal High Resolution Graphics
- Glitch, Pattern, Qualified, Interval, Dropout and Video Triggers
- 8-bit vertical resolution, 11 with ERES option
- DOS Compatible Floppy Disk, PCMCIA potable hard drive and Memory Card options
Memory Per Channel |
Channels Used |
Maximum Sample Rate
|
9350A 9354A |
9350AM 9354AM |
9354TM |
9350AL 9354AL |
Notes |
All,
Peak Detect OFF |
500 MS/s |
50k |
250k |
500k |
2M |
All channels active |
All,
Peak Detect ON |
100 MS/s data 400 MS/s peak |
25k data
+
25k peaks |
100 data
+
100 peaks |
250k data
+
250k peaks |
1M data
+
1M peaks |
All channels active 2.5 ns peak detect |
Paired Peak Detect OFF |
1 GS/s |
100 |
500 |
1M |
4M |
9350;
CH1 9354A;
Ch2 + Ch3 |
Paired + PP093 Peak Detect OFF |
2 GS/s |
200 |
1000 |
2M |
8M |
9354A Models Only |
采集系统
- 带宽(-3 dB):DC to 500 MHz
- No. of Channels:4 CH
- No. of Digitizers:4 CH
- 最大采样率:2 GS/s
- 采集存储器:9354AL可达8 M,9354TM可达2M.
- 灵敏度:2 mV/div to 5 V/div, fully variable.
- Scale factors:A wide choice of over 12 probe attenuation factors are
selectable.
- Offset Range:2.0 - 9.9 mV/div:± 120 mV
10.0 - 199 mV/div:± 1.2 V
0.2 - 5.0 V/div:± 24 V
±20 V across the whole sensitivity range when using the AP 020 FET probe.
- DC Accuracy:< or equal to±2% full scale.
- 垂直分解度:8 bits.
- 带宽限幅器:30 MHz
- 输入耦合:AC, DC, GND.
- 输入阻抗:1 M Ohm//14 pF or 50 Ohms ± 1%.
- 最大输入:
1 M Ohm:250 V (DC+peak AC < or equal to 10 kHz)
50 Ohms:± 5 V DC (500 mW) or 5 V RMS
时基系统
- Timebases:Main and up to 4 Zoom Traces.
- Time/Div Range:1 ns/div to 1000 s/div.
- Clock Accuracy:< or equal to 10 ppm
- Interpolator resolution:10 ps
- Roll Mode:Ranges 500 ms to 1,000 s/div.
- For > 50k points:10 s to 1,000 s/div.
- External Clock:< or equal to 100 MHz on EXT input with ECL, TTL or zero
crossing levels up to 500 MHz with option 935XA-CKTRIG.
触发系统
- Trigger Modes:Normal, Auto, Single.
- Trigger Sources:CH1, CH2, Line, Ext, Ext/10 (9354:CH3, CH4), Slope, Level and
Coupling for each can be set independently.
- Slope:Positive, Negative.
- Coupling:AC, DC, HF, LFREJ, HFREJ.
- Pre-trigger recording:0 to 100% of full scale (adjustable in 0.1 div
increments).
- Post-trigger delay:0 to 10,000 divisions (adjustable in 0.01% increments).
- Holdoff by time:10 ns to 20 s.
- Holdoff by events:0 to 99,999,999 events.
- Trigger Bandwidth:Up to 500 MHz using HF coupling.
- Internal Trigger Sensitivity Range:± 5 div.
- EXT Trigger Max Input:
1 M Ohm//15 pF:250 V (DC+peak AC < or equal to 10 kHz)
50 Ohms ± 1%:± 5 V DC (500 mW) or 5 V RMS
- EXT Trigger Range:±0.5 V (±5 V with Ext/10)
- Trigger Timing:Trigger Date and Time are listed in the Memory Status Menu.
SMART触发类型
- Pattern:Trigger on the logic AND of 5 inputs - CH1, CH2, CH3, CH4, and EXT
Trigger, (9350:3 inputs - CH1, CH2, EXT) where each source can be defined as
High, Low or Don't Care. The Trigger can be defined as the beginning or end of
the specified pattern.
- Signal or Pattern Width:Trigger on glitches as short as 1 nsec or on pulse
widths between two limits selectable from < 2.5ns to 20s.
- Signal or Pattern Interval:Trigger on an interval between two limits selectable
from 10ns to 20s.
- Dropout:Trigger if the input signal drops out for longer than a time-out from
25ns to 20s.
- State/Edge Qualified:Trigger on any source only if a given state (or
transition) has occurred on another source. The delay between these events can
be defined as a number of events on the trigger channel or as a time interval.
- TV:Allows selection of both line (up to 1500) and field number (up to 8) for
PAL, SECAM, NTSC or non-standard video.
采集模式
- Random Interleaved Sampling (RIS):for repetitive signals from 1 ns/div to 2 µs/div (M,L versions:from 1 ns/div to
5 µs/div).
- Single shot:for transient and repetitive signals from 10 ns/div (all channels
active).
- Peak detect:captures and displays 2.5 ns glitches or other high-speed events.
- Sequence:Stores multiple events - each of them time stamped - in segmented
acquisition memories.
- Number of segments available:9354TM 2-500, 9354AL 2-2,000
信号显示
- Waveform style:Vectors connect the individual sample points, which are
highlighted as dots. Vectors may be switched off.
- CRT:12.5 x 17.5 cm (9" diagonal) raster.
- Resolution:810 x 696 points.
- Modes:Normal, X-Y, Variable or Infinite Persistence.
- Real-time Clock:Date, hours, minutes, seconds.
- Graticules:Internally generated; separate intensity control for grids and
waveforms.
- Grids:1, 2 or 4 grids.
- Formats:YT, XY, and both together.
- Vertical Zoom:Up to 5x Vertical Expansion (50x with averaging, up to 40 µV
sensitivity).
- Horizontal Zoom Factors:9354TM 2,000x, 9354AL 40,000x
内置存储器
- Waveform Memory:Up to four 16-bit Memories (M1,M2,M3,M4).
- Processing Memory:Up to four 16-bit Waveform Processing Memories (A,B,C,D).
- Setup Memory:Four non-volatile memories. Optional Cards or Disks may also be
used for high-capacity waveform and setup storage.
自动测量
- Relative Time:Two cursors provide time measurements with resolution of ±0.05%
full scale for unexpanded traces; up to 10 % of the sampling interval for
expanded traces. The corresponding frequency value is also displayed.
- Relative Voltage:Two horizontal bars measure voltage differences up to ±0.2% of
fullscale in single-grid mode.
- Absolute Time:A cross hair marker measures time relative to the trigger, and
voltage with respect to ground.
- Pass/Fail testing allow up to five of the listed parameters to be tested against
selectable thresholds. Waveform Limit Testing is performed using templates which
may be defined inside the instrument.
波形分析
- Up to four processing functions may be performed simultaneously. Functions
available are:Add, Subtract, Multiply, Divide, Negate, Identity, Sin (x)/x and
Summation Averaging.
- Average:Summed averaging of up to 1,000 waveforms in the basic instrument. Up
to a million sweeps are possible with Option WP01.
- Envelope:Max, Min, or Max and Min values of up to one million sweeps.
- ERES:Low-Pass digital filter provides up to 11 bits vertical resolution.
Sampled data is always available, even when a trace is turned off. Any of the
above modes can be invoked without destroying the data.
- FFT:Spectral Analysis with four windowing functions and FFT averaging.
- Envelope and ERES modes are provided in Math Package WP01, FFT is in WP02.
自校
- Pressing Autosetup sets timebase, trigger and sensitivity to display a wide
range of repetitive signals. (Amplitude 2mV to 40V; frequency above 50Hz; Duty
cycle greater than 0.1%).
- Autosetup Time:Approximately 2 seconds.
- Vertical Find:Automatically sets sensitivity and offset.
界面连接
- Remote Control:All front-panel controls, as well as all internal functions are
possible by GPIB and RS-232-C.
- RS-232-C Port (Standard):Asynchronous up to 19200 baud for computer/terminal
control or printer/plotter connection.
- GPIB Port (Standard):(IEEE-488.1) Configurable as talker/listener for computer
control and fast data transfer. Command Language complies with requirements of
IEEE-488.2.
- Centronics Port:Optional hardcopy parallel interface included with floppy disk
and graphics printer option.
- Hardcopy:Screen dumps are activated by a front-panel button or via
remote control. TIFF format is available for importing into Desktop
Publishing programs. The following printers and plotters can be used to make
hardcopies:
HP DeskJet (color or B&W), HP ThinkJet, QuietJet, LaserJet, PaintJet and EPSON
printers. HP 7400 and 7500 series, or HPGL compatible plotters.
An optional internal high resolution graphics printer is also available.
概要
Auto-calibration ensures specified DC and timing accuracy.
- Temperature:5° to 40° C (41° to 104° F) rated, 0° to 50° C (32° to 122° F)
operating.
- Humidity:< 80%.
- Shock & Vibration:Meets MIL-STD-810C modified to LeCroy design specifications
and MIL-T-28800C.
- Power:90-250 V AC, 45-66 Hz, 230 W.
High speed and long memory make this family the ideal 500 MHz general-purpose
Digital Storage Oscilloscopes. Two and four channel simultaneous sampling at 500
MS/s meets demanding high-speed design applications. Even faster sampling may be
achieved by combining channels, up to a maximum of 2 GS/s. Acquisition memories
may also be combined, providing up to 8 M points of continuous or segmented
waveform recording. Repetitive signals are digitized at up to 10 GS/s.
A unique peak detect scheme triggers on glitches down to 1ns and keeps the
ADC sampling at 2.5 ns - even at slow time bases - without destroying the
underlying data. This provides circuit designers with the benefits of peak
detection without any loss of precision.
Live waveforms on the main timebase may be viewed simultaneously with up to 3
expansions, showing all of the signal detail. Expansions are shown as highlights
on the main trace. Each zoomed detail may be expanded horizontally and
vertically as required.
SMART Trigger modes like Glitch, Pattern, Dropout and TV allow you to capture
precisely the events of interest. Pre- and Post-Trigger delay, and Time and
Events Holdoff are also standard. The 9350A family features the proven
user-interface of LeCroy's portable scopes. A bright 9" CRT allows optimum
waveform viewing on a high resolution 810 x 696 pixel screen. Menus and text are
arranged around the graticules - they never overwrite the waveforms. Dedicated
control knobs keep the scope's performance at your fingertips.
A comprehensive range of signal processing functions including FFT and Math
on live or stored waveforms, allows extensive waveform manipulation. Up to 16
MBytes of RAM are standard allowing advanced processing including FFT's up to 1
Mpoint. For the most powerful processing in the industry, a 64 MByte RAM option
is available. DOS compatible floppy disk, PCMCIA portable hard drive and memory
card options store waveforms and test setups, and simplify data transfer to any
PC. An optional high resolution graphics printer is also available.
咨询电话: 132- 4173- 5460 , 邮箱: 52634729@QQ.com